    MDCG 2019-16 Rev.2 at six months: Notified Body audit patterns

    Six months in, Notified Body audits under MDCG 2019-16 Rev.2 show three clear failure modes - narrative-only SBOMs, CVD policies without operational evidence, and traceability gaps between security risk controls and design outputs.

    Reviewed by ·Published ·Last reviewed July 2026

    Six months after MDCG 2019-16 Rev.2 (April 2026) landed, Notified Body audit findings under it cluster around three failure modes:

    1. SBOMs described in prose but not delivered as machine-readable CycloneDX or SPDX files. The Rev.2 expectation is explicit - narrative alone is a nonconformity.

    2. Published CVD policies without operational evidence. Auditors are asking for redacted example cases, triage SLA definitions, and evidence that a report has ever been processed end-to-end.

    3. Traceability gaps between security risk control measures (in the security risk file) and design outputs (in the DHF/technical file). Rev.2 expects a matrix, not a narrative.

    Action for MDR holders: if you passed your last surveillance audit under Rev.1 language, don't assume Rev.2 posture. Rerun a gap review this quarter.