Six months after MDCG 2019-16 Rev.2 (April 2026) landed, Notified Body audit findings under it cluster around three failure modes:
1. SBOMs described in prose but not delivered as machine-readable CycloneDX or SPDX files. The Rev.2 expectation is explicit - narrative alone is a nonconformity.
2. Published CVD policies without operational evidence. Auditors are asking for redacted example cases, triage SLA definitions, and evidence that a report has ever been processed end-to-end.
3. Traceability gaps between security risk control measures (in the security risk file) and design outputs (in the DHF/technical file). Rev.2 expects a matrix, not a narrative.
Action for MDR holders: if you passed your last surveillance audit under Rev.1 language, don't assume Rev.2 posture. Rerun a gap review this quarter.
- Three years of Section 524B: the deficiencies that never went awayThree-year retrospective on Section 524B enforcement: SBOM depth, VEX justification, and CVD operational evidence remain the top three categories of cyber deficiencies. Two new patterns emerged in 2026 - AI/ML threat-model gaps and weak end-of-support labeling.
- IEC 81001-5-1 Amendment 1 reaches FDIS - what MedTech teams should trackAmendment 1 to IEC 81001-5-1 (health software security lifecycle) is in Final Draft International Standard ballot. It sharpens SBOM, CVD, and postmarket vulnerability expectations - and it's the standard EU Notified Bodies are quietly aligning MDCG 2019-16 audits against.
- EU CRA delegated acts under Article 27 - the technical baseline is taking shapeThe European Commission's Article 27 delegated acts translating the Cyber Resilience Act's essential requirements into technical standards are moving through consultation. For medical device makers, three of the draft harmonized standards will do most of the audit work.